Hightemperatureagingbox高温老化箱、高温老化试验箱
Equipment uses: high temperature aging box for electronic and electrical, automotive, aerospace, rubber, plastic, metal, ship weapons. Under the condition of high temperature and constant temperature change of the parts and materials of the related products, such as institutions of higher learning and scientific research, the performance indexes of the products are tested. Strictly according to GB/T2423.2-2008 high temperature aging test method.
Main technical parameters of the equipment specification material:
model(L) | Studio size寸D*W*H(mm) | Carton sizeD*W*H(mm) | Total power(K.W) | temperature range |
TY/GW-50L | 350*350*400 | 790*800*1200 | 3 |
RT+10~100℃
RT+10~200℃ RT+10~250℃ RT +10~300℃ |
TY/GW-100L | 400*500*500 | 970*850*1300 | ||
TY/GW-150L | 500*500*600 | 970*950*1350 | 5 | |
TY/GW-225L | 500*600*750 | 1090*1130*1700 | ||
TY/GW-400L | 600*800*850 | 1170*1300*1750 | 6 | |
TY/GW-500L | 700*800*900 | 1182*1250*1800 | ||
TY/GW-800L | 800*1000*1000 | 1250*1500*1850 | 9 | |
TY/GW-1000L | 1000*1000*1000 | 1450*1500*1850 | ||
Note: other temperature size of the material can be customized according to requirements, equipment in kind prevail, the experimental data are in 25 DEG C no-load conditions measured | ||||
Construction | ||||
Internal material | 304SUS stainless steel plate | |||
External material | High quality cold-rolled steel electrostatic spraying, strong, beautiful, durable, durable, color optional | |||
material | Superfine glass wool, or foaming material, high temperature insulation performance is excellent | |||
heater | Stainless steel, nickel chromium alloy fin type heater | |||
Sealing silicone | Anti aging high temperature resistant silicone high temperature of 400 DEG C | |||
Air duct system | ||||
fan | Fan for long shaft | |||
Yi Ting instrument control system | ||||
Control system | Japan's Fuji key type intelligent instrument setting operation mode has the advantages of simple operation, direct heating plus or minus button shows the actual temperature and the set temperature Touch screen optional TEMI series of procedures or fixed value of two kinds of control mode, can be connected to the computer, in English switching. | |||
Electrical components | ||||
Electrical | Selection of LG/ / CHINT / Schneider / West Germany and other well-known brands | |||
other | ||||
Power Supply | AC220V~AC380V±10%50Hz±0.5Hz | |||
safeguard | Over temperature alarm, leakage protection, under phase phase and phase lack protection, over current protection, fast fuse, line fuse and all sheath type terminal, grounding protection | |||
Meet the standard | GB2423.2-2008 (IEC68-2-2) test B: high temperature test method SJT31391-1994 test requirements of readiness and methods of inspection and assessment |
(model appearance of technical information are subject to change without notice, Ting Yi has the power of final interpretation)
Yi Ting high temperature test box is provided with strong high price | |||
| |||
A large screen high performance controller three yuan /TEMI series Powerful functions, easy to use, interface pretty outrageous | Using the computer technology, the advanced PID control method to measure and control the temperature and humidity of the environment, and can be used with constant temperature and humidity equipment. Alternating high and low temperature test box, suitable for more suitable test box with constant temperature and humidity, dust and rain xenon lamp UV ozone salt fog test box. Can display and control interface is clear and intuitive touch, select menu, easy to use, stable and reliable performance. This program control is flexible, cost-effective. | ||
True color touch screen 3.5 inch /4.3 inch /5.7 inch and 7 inch /9 inch Two 2 control methods: program / set value, the program can edit 120 sets of programs, the maximum can be compiled for 100 sections, the group of all cycle, the segment cycle; 3 sensor type: two PT100 input (optional electronic sensor input 4 products show clear intuitive three-dimensional sense of strong, programmable control system is flexible and convenient operation, stable performance, more efficient work; flexible installation method, 5 can be embedded or embedded installation 6 alarm signal: 16 DI external obstacle alarm; 7 temperature measurement range: --200.00 C -90.00 C, (optional --300.00 C -90.00 C) error + 0.2 c; 8 humidity measurement range: 1.0%--100%RH, the error is + 1%RH; 9 communication interface: (RS232/RS485, communication the longest distance 1.2km[optical fiber up to 30km]); 10 interface language type: Chinese / english; 11 Chinese character input function, edit input manufacturers information, fault name, test name, etc., display intuitive, clear; 12 curves: real time temperature, humidity PV, SP curve; 13 with network function IP address can be set; can bring the printer (USB function optional); remote control instrument; | |||
Two. / Schneider / CHINT and other well-known Delixi electrical components Performance is stable and unparalleled, long and durable to prevent the problem of circuit | |||
Three, the Ni Cr alloy fin type heating The embedded heating element is arranged in the stainless steel cavity interlayer, which is extremely safe, fast temperature rise, no oxidation, and long service life. | |||
Four, safe and secure, a variety of protection systems do not have to worry about hidden dangers Equipped with over temperature alarm, leakage protection, under phase phase and phase lack protection, over current protection, fast fuse, line fuse and full jacket type terminal, water shortage protection, grounding protection. |
AC220v/380V, three-phase four wire + ground protection; In the frequency range: 50Hz + 0.5 Hz Protection, earthing resistance is less than 4 TN-S or TT power supply; power supply mode, The protective grounding wire, separate grounding, grounding wire and other equipment not shared. |
○ The use of ambient temperature 5 to 35 DEG C, the best use temperature: 15 to 25 DEG C, relative humidity: atmospheric pressure: 86-106Kpa = 85%RH |
In ground level, good ventilation, no flammable, explosive and corrosive gases and dust No strong airflow, around when the surrounding air need forced flow, air flow should not blow directly into the box No, no electromagnetic field near direct sunlight and other thermal radiation source In the peripheral equipment have adequate maintenance space Ground, bearing requirements more than 600kg/ square meters (the upstairs venue) |
Do not force the glass, transportation, do not impact the body. , the equipment should be placed on a smooth landing, such as uneven ground should be amended. This equipment, in normal operation, the placement of the material shall not influence the air circulation to ensure uniform temperature in the box. Behind, don't close to the power line, do not put the instrument or other materials on the power line, so as not to damage the power line。 |
Maintenance considerations and methods:
1, flammable and volatile chemicals do not put into the box.
2, such as the use of abnormal, odor, smoke, etc., please immediately shut off the power supply, the user must not blindly repair, we should inform the repair department, the professional personnel to see repair.
3, the inner of the cabinet and the equipment always wipe the surface, to keep clean, to increase the transparency of the glass. Do not use acid, alkali or other corrosive solutions to wipe the outer surface.
4, equipment for a long time not, should unplug the power cord to prevent damage to the equipment. And should be regularly (usually a quarter) according to the conditions of use 2-3 days, to get rid of the moisture in the electrical parts, to avoid damage to the device.
troubleshooting
phenomenon | principle | handle |
1 no power | 1 no power outlet | 1 switch socket |
2 plug is not inserted or broken | 2 insert the plug or connect the wire | |
3 fuse open circuit | 3 switch fuse | |
4 the power switch is not closed. | 4 close power switch | |
2 the temperature in the box is not increased. | 1 set the temperature is low | 1 adjust the set temperature |
2 electric heater is bad | 2 electric heater | |
3 temperature controller is bad | 3 change control Wen Yi | |
4 temperature sensor connection line loose | 4 tighten the connecting wire nut of the sensor | |
3 set temperature and the temperature of the large error | 1 temperature sensor is bad | 1 Change |
2 ambient temperature is high (more than 30 degrees Celsius) | 2 put in an air-conditioned environment | |
4 temperature out of control | 1 temperature sensor fixed off | 1 fixed temperature sensor |
2 temperature control instrument is bad, controllable silicon is bad | 2 for temperature control, SCR |