Ting Yi talk about how to maintenance the test box
First, adhere to professional management, conditional units should not regularly send someone to the supplier's factory training to learn. In order to obtain a more professional maintenance, repair experience and ability.
Second, recommendations will be placed in relatively benign environment temperature, optimum temperature and humidity of the environment: 5 DEG C to being DEG C is less than or equal to 85%R.H, do not have the conditions of laboratory, shall be equipped with appropriate air conditioner or cooling tower. In order to ensure the service life of the compressor and the machine's overall results.
Third, circulating fan, condenser fan clean and balance, and cleaning evaporator similar, because of the test box in the working environment of different, circulating fan and condenser fan will be condensed a lot of dust and other small particles objects, should be regularly cleaned.
Four, water and humidifier cleaning: if water is not smooth, humidifier scaling easily lead to humidifier dry burning, may damage the humidifier, so you must clean the water and humidifier regularly.
Regular equipment clean, and ensure performance of heat exchanging of a good ventilation, for compressor adopts water cooling, besides must guarantee its inlet pressure and inlet temperature must also be ensure the corresponding flow. And regularly on internal condenser were cleaning to obtain the sustained heat transfer performance.
Six, if need to remove is preferably carried out in the court tomorrow instrument technical staff under the guidance, so as to avoid unnecessary damage or refrigerant leak.
Seven, the long term does not use should be clean and clean, keep dry, every half a day to the device through a power, time is not less than 1 hours